C alib ratio n --T yp e IL30
* c. Push forw a rd on the dial tape, to free the tape from
thes sprocket, end position the tape to read 1750.
2. Calibrate Low Frequency End
a.
Tune
the
frequency
meter to 650 M Hz.
(The
2nd har
monic will
be used
to set
the
oscillator frequency.)
b.
Tune the oscillator to read 1100 on the dial tape.
c. Adjust the varactor until the oscillator frequency is
1300 M Hz as indicated by a beat mode signal on the fre
quency meter.
Turning the Varactor assembly clockwise will reduce the
oscillator frequency. Turning the Varactor assembly counter
clockwise will increase the oscillator frequency..
3. Repeat steps 1 and 2 as necessary to track
the oscillator to the tape.
4. Check Low Frequency Mechanical Stop
a. Tune the oscillator td the low frequency mechanical
stop.
I*--''
fa
ir
U:
&
f
i:
i-V.
i
f.
I
$■
t v
S' ■
I"
B
©
6 -3 2