Fig. 4-6. Spanner wrench required in the removal of the oscillator
tube. Tektronix Part No. 003-0397-00.
10. Recalibrate as directed in the Calibration procedure.
Replacing the M ixer Diode
1. Remove the front panel mounting nut and washer for
the RF INPUT connector.
2. Slip the mixer assembly back and out of the unit.
3. Unscrew the barrel (1 dB pad) (see Fig. 4-7) and re
place the mixer diode.
M a in te n a n c e — Type 1L30
Removing and
Remounting the H oneycom b
Assembly
1. Loosen the front set-screw on the coupler to the DIS
PERSION RANGE switch shaft with an Allen wrench. Slide
the shaft out through the front panel so it is out of the way.
2. Remove
the two screws securing
the DISPERSION
RANGE switch mounting hardware to the rear plate and
swing the switch assembly out of the w ay behind the rear
plate.
3. Remove the fourteen Phillips head screws fastening the
IF chassis to the base assembly (top plate).
4. Swing the chassis up and out, to rest on the instru
ments spacer bars (see Fig. 4-8). It may be necessary to
disconnect the coaxial cable from J147. Do not use force,
as some of the parts are critically positioned and should
not be moved out of adjustment.
5. Insure that none of the terminals and tie points are
shorted or grounded. Reconnect any cables or wires that
may be disconnected. Fig. 4-9 illustrates wiring color code.
Apply power if desired. The ground-wire on the DISPERSION
RANGE switch must be grounded for proper operation of
the instrument. Fig. 4-15 illustrates the component layout and
circuit layout on the honeycomb chassis.
6. Remount the chassis using the reverse procedure of
steps 1 through 5. Do not force the chassis into place.
Check for pinched or undue strain on the wires and connec
tors. When replacing the DISPERSION RANGE switch, make
certain its shaft is properly coupled to SW365. Check the
operation of the DISPERSION RANGE switch and insure that
the knob is properly indexed with the front panel markings.
W ire and cable color code are shown in Fig. 4-9.
Fig. 4-7. location of th® mixer diode.
Removing
the Phase Lock
Assembly
1. Switch the power to OFF and remove the instrument
from the plug-in oscilloscope.
2. Unplug the. signal lead (J855). from the phase-lock
assembly.
3. Loosen the set screw for the FINE RF CENTER FREQ
control and remove the knob.
4. Use a
inch nut driver to remove the mounting nuts
securing the front panel phase-lock controls (FINE RF FREQ,
LOCK CHECK and INT 1 M H z REF FREQ switch). Keep the
nut for the INT 1 M Hz REF FREQ switch separate from the
others, because it has a different thread and will bind if it
is placed on the wrong control.
C A U T IO N
Do not loosen or move the pickup probe in the
oscillator assembly. Its position is critical for pro per
o peratio n of the oscillator.
5. Remove the six Phillip head screws located along the
edge of the "U" shaped phase-lock assembly cover.
4-5
I
r
i:
f
I
£
h
%
ir
r
V
I
©
Summary of Contents for 1L30
Page 48: ...si L i y n Maintenance Type 1L30 Fig 4 15 Honeycomb assembly drcui a n d component layout 4 n ...
Page 59: ...Fig 6 1 A Test equipment required for calibration ...
Page 60: ... Calibration Type 1130 ...
Page 120: ...T Y P E I L 3 0 S P E C T R U M A N A L Y Z E R ib i IF SYSTEM BLOCK DIAGRAM 9 1 9 2 ...
Page 127: ......