![Tektronix 1L30 Instruction Manual Download Page 14](http://html1.mh-extra.com/html/tektronix/1l30/1l30_instruction-manual_1077897014.webp)
Operating instructions— Type 1L30
■
■
I
&
I
i$
3
■1
:4
i
turn the RESOLUTION control to the coupled position with
the DISPERSION selector.
GENERAL OPERATING INFORMATION
Signal Application
Signals applied to the RF INPUT connector should be
applied through a 50-ohm coaxial cable with a N type
connector. Impedance mismatches between the signal source
and the RF INPUT connector, due to signal source output
impedance, long coaxial cables, due to signal source output
affect on display flatness. When optimum flatness is de
sired and signal strength is adequate, a 50 ft attenuator pad
(such as 10 dB pad) should be added between the signal
source and the RF INPUT connector.
This will minimize
the reflections and optimize display flatness.
Signal input power to the analyzer should not exceed
—30 dBm. Signals above this level may overload the 1st
mixer and/or the 1st IF stage and generate spurious sig
nals on the display. Add at least 10 dB attenuation to the
input when signal compression is noticed (no increase in
signal height with an increase in input power). A conver-.
sion chart (Fig. 2-6) may be used to calculate input signal
levels.
C A U T IO N
Signals whose amplitudes are g rea ter than + 1 5
dBm ap p lied to the INPUT may d am a g e the diode
mixers.
Front Panel Calibration Adjustments
Three screwdriver adjustments provide a means to cali
brate the DISPERSION, IF CENTER FREQ and the DISPER
SIO N RANGE balance. These front panel adjustments must
be recalibrated whenever the Type 1L30 is shifted to an
other oscilloscope, to compensate for differences in sawtooth
amplitudes and CRT deflection sensitivities. It is also ad
visable to check the adjustments periodically during regu
lar use. Adjustment and a calibration check procedure may
be performed as follows:
1. Balance and Calibration Check
a. Set the INT 1 M Hz REF FREQ switch in the OFF posi
tion, then tune a signal on screen with the RF CENTER FREQ
control.
b. Tune for minimum signal shift as the DISPERSION
RANGE is switched from M H z /C M to kH z/C M positions.
c. W ith the DISPERSION RANGE selector at the M H z /C M
position, adjust the IF CENTER FREQ control for minimum •
signal' shift as the DISPERSION selector is switched through
the 10 to .2 M Hz positions.
d. Set the DISPERSION RANGE selector to kH z/C M posi
tion. Continue adjusting the IF CENTER FREQ control for
minimum signal shift as the DISPERSION is switched from
500kHz/cm to 5k H z/cm .
e. Set the DISPERSION to the 500 kHz/cm position, then
center the signal in the graticule area with the Horizontal
Position control. Check the signal position on the sweep.
The signal should locate within ± 1 cm of the sweep center
Fig. 2-6. Volts-dBm-Wafts conversion chart for 50 ft impedance.
2 - 8
©