![Tektronix 1L30 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/tektronix/1l30/1l30_instruction-manual_1077897046.webp)
i-iw i Jvieei iti*
Maintenance— Type :■: .10
ode with the typical voltages listed on the diagram will help
isolate faulty, diodes.
Forward-to-back resistance ratios
on some diodes can be checked by referring to the schematic
and pulling appropriate transistors and square pin connectors
to remove low resistance loops around the diode.
C A U T IO N
Do not use an ohm m eter scale th at has a high in
ternal current. Do not check the fo fw a rd -to -b a c k
resistance ratios of tunnel diodes or mixer diodes.
Some Trouble Symptoms
A misleading trouble symptom may occur if
one o f the
Varactor diodes in the oscillator circuit is shorted.
This will
clamp the DC output voltage from the
phase lock circuit and
prevent vertical trace shift as the FINE RF CENTER
FREQ
con
trol is adjusted. The symptom indicates trouble in the
phase
lock circuit when it is actually in the tuner.