Fig. 6-27. Typical equipment setup to check sensitivity,
Type 1L30
POS'
Position a free running
trace to the bottom
line of the graticule.
IF CENTER FREQ
Midrange (000)
FINE
Centered
DISPERSION RANG E
k H z /C M
DISPERSION
5 0 0 kHz
VIDEO FILTER
OFF
G A IN
Midrange
VERTICAL DISPLAY
LIN
IF ATTEN
Switches off
FINE RF CENTER FREQ
Centered
IN T 1
M Hz
REF FREQ
OFF
Oscilloscope
Tim e/Cm
2 0 ms
Triggering
Adjusted for a free
running sweep
Horizontal Position
Centered trace
dial accuracy and phase lock operation (step 1 8 ).
18. Check RF Center Frequency Calibration,
System Sensitivity and Phase Lock
a. Equipment setup is shown in Fig. 6-27.
t
\
b. Apply a frequency and amplitude calibrated signal
;#
that is between — 60 dBm and — 30 dBm to the RF INPUT
£
I
If an external atte n u ato r is used, it must have fla t
high frequency characteristics: Use Tektronix 2 0
dB A tte n u ato r Part N o . 0 1 1 - 0 0 8 6 - 0 0 , or
A tte n u ato r Part N o . 0 1 1 - 0 0 8 7 - 0 0 .
c. Set the DISPERSION control to 500kHz/cm and the
RESOLUTION control for a resolution bandwidth of 100
kHz (fully clockwise).
d. Adjust the G A IN control for an average noise ampli-
f
tude of one centimeter. Center the IF CENTER FREQ and
k
FINE RF CENTER FREQ controls.
I.
e. Tune the signal on screen with the RF CENTER FREQ
(■„
control. Reduce the signal amplitude with the signal gener-
t
'
©
6 -2 7