Cafii:
ire ii c n— Type 1L30
Fig. 6-3. Equipment setup to
SWEEP CIRCUIT
Type
1L30
POS
Position a free
trace to the
line of the g
IF CENTER FREQ
Midrange (000)
FINE
Centered
CAL (Adjustment)
Centered
DIS CAL
(Adjustment)
Centered
DISPERSION RANGE
M H z /C M
DISPERSION •
10 M Hz
VIDEO FILTER
OFF
G A IN
Midrange
VERTICAL DISPLAY
LIN
IF ATTEN
Switches off
FINE RF CENTER FREQ
Centered
IN T 1 M H z REF
FREQ
OFF
MIXER PEAKING
Fully ccw
Time/Cm
Triggering
Horizontal Position
Oscilloscope
10 ms
Adjusted for a free
running sweep
Centered trace
1. Adjust Center Frequency
O
a. Equipment setup is shown in Fig. 6-3.
b. Connect the probe of a DC coupled test oscilloscope
to chassis ground on the Type 1L30. Establish 0 V reference
level on the test oscilloscope, then connect the probe to
pin P of the square pin connector for the honeycomb as
sembly.
c. Adjust the IF CF Range R290 (see Fig. 6-4) for 0.75 volts
± 0 .1 V of trace deflection on the test oscilloscope.
»
d. Disconnect the probe from pin P of the connector.
e. Apply a 200 M Hz signal from the Time-Mark G enera
tor (harmonic of 10 ns marker) through a 20 dB attenuator
to the RF INPUT connector.
■
f. Adjust the G A IN control for a display signal ampli
tude of approximately 4 cm. Tune the RF CENTER FREQ
control to minimize converted signal interference. The 200
M Hz IF feed-through signal should be relatively free of
spurious signal interference.
g. Adjust the Center Freq Cal R253 (Fig. 6-4) for mini
mum IF signal shift as the DISPERSION selector is switched
between the 10 M Hz and .2 M Hz positions. The DISPER
SIO N RANGE switch must be in the M H z /C M position for
this adjustment.
h. Return the DISPERSION selector to the 10 M Hz posi
tion and adjust the Horizontal Position control to center
the sweep on the graticule.
6 -7
Summary of Contents for 1L30
Page 48: ...si L i y n Maintenance Type 1L30 Fig 4 15 Honeycomb assembly drcui a n d component layout 4 n ...
Page 59: ...Fig 6 1 A Test equipment required for calibration ...
Page 60: ... Calibration Type 1130 ...
Page 120: ...T Y P E I L 3 0 S P E C T R U M A N A L Y Z E R ib i IF SYSTEM BLOCK DIAGRAM 9 1 9 2 ...
Page 127: ......