10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • [email protected]
7
GENERAL INFORMATION
WARNING
The American National Standards Institute (ANSI) states that
a shock hazard exists when probes or sensors are exposed
to voltage levels greater than 42VDC or 42V peak AC. Do
not use this instrument where voltages at the measurement
surface exceed these levels.
WARNING
The battery is accessible through a cover on the back of the
instrument. To avoid electrical shock hazard, disconnect all
temperature probes and sensors and turn the unit off before
removing the cover.
WARNING
Never use this instrument or any probe or sensor inside a
microwave oven.
CAUTION
Avoid making sharp bends in probe or sensor lead wires.
Bending lead wires at a sharp angle can damage the wire
causing probe failure.
CAUTION
Keep inputs electrically isolated from outputs. Do not
connect an external voltage source between the two
connectors, and avoid any external electrical pathways
between input and output (T2 & T1).
Failure to do so will disturb meter readings and calibrator
output levels. Damage to the instrument is also possible.
Where the meter and calibrator connectors are
simultaneously wired into a system lacking the above
isolation, it is necessary to use an "ungrounded" (i.e.
electrically isolated) probe at the thermometer input.