SERVICE INFORMATION
the same temperature should result in a null, or 0.000 mV on the
voltmeter.
11. Disconnect the cables from the equipment and UUT and reconfigure the set-up per figure 12.
12. Select the meter mode by pressing the CALIB/METER function button on the U.U.T. and
ensure the U.U.T. is still set for °F and type “K” thermocouple.
13. Set the thermocouple simulator to type “K” and output an output of 32.0°F. The U.U.T.
should read 32.0°F.
14. Set the thermocouple simulator to 2500°F. The U.U.T. should read 2500°F.
15. Set the thermocouple simulator to - 320°F. The U.U.T. should read -320°F.
16. Remove the Alloy Mode Calibrator Verification Cable from all equipment.
17. The Alloy Mode verification is now complete. Press the OFF function button to turn the unit
off.