SERVICE INFORMATION
thermocouple by pressing the sensor/select button. The voltmeter should read 0.000mV.
40. Press the OFF function button. The unit will turn off.
41. Press the ON function button. The unit will power up and return to the calibrator mode.
"CALIB" will appear on the display along with the "K" enunciator, the "OPR" enunciator, and
"32.0 °F" will be displayed. The blinking "Ω" symbol will have completely disappeared
indicating that you have exited the copper mode.
42. Remove the Copper Mode Calibrator Verification Cable from both the voltmeter and the
TEGAM UUT.
43. The copper mode verification is now complete. Press the OFF function button to turn the unit
off.
ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS
Equipment Required:
1. TEGAM Model to be verified.
2. Precision temperature simulator / voltage source.
3. Voltmeter accurate to 1 uV resolution.
4. Cable assembly.
1. Start your calibration with a fresh battery.
2. Press the ON function button to turn the unit on.
3. Select the calibrator mode ("CALIB" on display) by pressing the CALIB/METER function
button.
4. Select type K thermocouple ("K" on display) by pressing the SENSOR SELECT function button
until type K is reached.
5. Select the tenth degree display with a digit appearing after the decimal point on the display)
by pressing the 0.0.1 function button.
6. Select °F ("°F" on display) by pressing the °F °C function button.
7. Attach the Alloy Mode Calibrator Verification Cables per Fig.13.
8. Press the CHANGE function button and the numerical display will blink and "NUM" will appear
on the display.
9. Press 3 2 0 buttons and then press the ENTER function button. "NUM" will disappear and
"32.0 °F" will appear, not blinking.
10. Adjust the Thermocouple Temperature Simulator to 32.0 °F.
NOTE: The reading on the voltmeter should read null. 0.000 mV. Adjusting the
U.U.T. to any temperature and adjusting the Temperature Simulator to