SERVICE INFORMATION
Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence that follows.
This will ensure that adjustments to be stored in EEPROM go to the
correct locations.
1. Set meter-mode zero and gain adjustments per Table 1.
2. Press OPR key (RTD annunciator turns off).
3. Set U.U.T. to CALIB, OPR, K, °F.
4. Turn U.U.T. OFF, then ON, to re-enable RTD & Ω, annunciators.
5. Set calibrator-mode zero and gain adjustments per Table 2. Press CHANGE and ENTER at the
completion of each step in Table 2.
6. Press OPR key (RTD annunciator turns off).
7. Install J1 jumper.
8. Set U.U.T. to METER, OPR, T, °F
9. Change TC simulator output to Type T, 32°F, ALLOY mode.
10. Reconfigure calibration set-up per Figure 12. Allow 2-3 minutes for thermal stabilization.
11. Calculate E1 as follows: E1 = U.U.T. Reading — 32.0.
12. Set U.U.T. to CALIB, OPR, T, °F mode.
13. Store E1; CHANGE/(E1 value)/ENTER/STO/1/OPR.
14. Go to METER mode, and verify U.U.T. reading of 32.0 ±0.1°F.
15. Reconfigure calibration setup per Figure 13. Leave simulator output at 32°F, Type T, ALLOY
mode. Allow 2-3 minutes for thermal stabilization.
16. Set U.U.T. to CALIB mode, 32°F, Type T.
17. Calculate E2 as follows:
E2 = DMM Reading (in μV)
-22
(For example with DMM reading of +11μV, E2 = -0.5°F)
18. Store E2; CHANGE/(E2 value)/ENTER/STO/2/OPR.
19. Install J2 jumper.
20. Re-set U.U.T. output to 32°F, DMM reading should be 0±2μV.
21. Turn OFF U.U.T.
22. Thermocouple calibration is complete. Remove calibration cover. Reinstall original back-cover,
unless going on to Part B (Resistance Calibration).