SERVICE INFORMATION
25. Press the SENSOR SELECT function button. The sensor type will change to type R. The "B"
will disappear from the display and a "R" will appear on the display.
26. Press the CHANGE function button. The "1000.0 °F" display will blink and "NUM" will appear
on the display.
27. Press 3 2 0 buttons and then press ENTER function button. "NUM" will disappear and "32.0
°F" will appear, not blinking. The voltmeter should read 0.000 mV.
28. Press the SENSOR SELECT function button. The sensor type will change to type S. The "R"
will disappear from the display and a "S" will appear on the display. The voltmeter should read
0.000 mV.
29. Press the SENSOR SELECT function button. The sensor type will change to type G. The "R"
will disappear from the display and a "g" will appear momentarily in the center of the display.
"LO" will then appear in the center of the display and both the "T" and "R" will appear along
the top row of the display.
30. Press the CHANGE function button. The "32.0 °F" display will blink and "NUM" will appear on
the display.
31. Press 6 0 0 0 buttons and then press ENTER function button. "NUM" will disappear and
"600.0 °F" will appear, not blinking. The voltmeter should read 2.238 mV.
32. Press the SENSOR SELECT function button. The sensor type will change to type C. A "C" will
appear momentarily in the center of the display.
"600.0" will then appear in the center of the display and both the "T" and "R" will appear along
the top row of the display.
33. Press the CHANGE function button. The "600.0 °F" display will blink and "NUM" will appear
on the display.
34. Press 3 2 0 buttons and then press ENTER function button. "NUM" will disappear and "32.0
°F" will appear, not blinking. The voltmeter should read 0.000 mV.
35. Press the SENSOR SELECT function button. The sensor type will change to type D. A "d" will
appear momentarily in the center of the display. "32.0" will then appear in the center of the
display and both the "T" and "R" will appear along the top row of the display. The voltmeter
should read 0.000 mV.
36. Press the SENSOR SELECT function button. The sensor type will change to type K. The "T"
and the "R" will disappear from the display and a "K" will appear on the display. The voltmeter
should read 0.000 mV.
37. Select the meter mode by pressing the CALIB/METER function button.
38. Output the voltages from table 1 into the U.U.T. from the thermocouple simulator, changing
the thermocouple type on the U.U.T. as noted in the table. Observe the correct readings on the
display of the U.U.T.
39. Select the calibrator mode by pressing the CALIB/METER function button and type “K”