SERVICE INFORMATION
Table 1: Meter-Mode Zero & Gain Adjustments
STEP
SENSOR SELECT U.U.T. INPUT
ADJUST
U.U.T. READING*
1
K
0.000mV
P1
32.0±0.1
°
F
2
K
54.856mV
P11
2500.0±0.1
°
F
3
Repeat steps 1 & 2 as required
4
J
0.000mV
P1
32.0±0.1
°
F
5
T
0.000mV
P1
32.0±0.1
°
F
6
E
0.000mV
P1
32.0±0.1
°
F
7
N
0.000mV
P1
32.0±0.1
°
F
8
B
1.439mV
P1
1000.0±0.5
°
F
9
R
0.000mV
P1
32.0±0.5
°
F
10
S
0.000mV
P1
32.0±0.5
°
F
11
G
2.238mV
P1
600.0±0.2
°
F
12
C
0.000mV
P1
32.0±0.2
°
F
13
D
0.000mV
P1
32.0±0.2
°
F
*exclusive of noise
Table 2: Calibrator-Mode Zero & Gain Adjustments
STEP
SENSOR SELECT U.U.T. INPUT
ADJUST
DMM READING*
1
K
32.0
°
F
P1
0.000mV±1µV
2
K
2500.0
°
F
P41
54.856mV±1µV
3
Repeat steps 1 & 2 as required
4
J
32.0
°
F
P1
0.000mV±1µV
5
T
32.0
°
F
P1
0.000mV±1µV
6
E
32.0
°
F
P1
0.000mV±1µV
7
N
32.0
°
F
P1
0.000mV±1µV
8
B
1000.0
°
F
P1
1.439mV±1µV
9
R
32.0
°
F
P1
0.000mV±1µV
10
S
32.0
°
F
P1
0.000mV±1µV
11
G
600.0
°
F
P1
2.238mV±1µV
12
C
32.0
°
F
P1
0.000mV±1µV
13
D
32.0
°
F
P1
0.000mV±1µV
*exclusive of noise