10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • [email protected]
3
GENERAL INFORMATION
DISPLAY:
5 digit LCD, 0.4" height, polarity indication and decimal point. Annunciators;
• Loaded Memory Locations (10 for data storage, 4 for ramp parameters)
• Low Battery (BAT)
• Number Change Mode (NUM)
• Calibrator Mode (CALIB)
• Meter Mode (METER)
• Operate Mode (OPR)
• Store Mode (STO)
• Recall Mode (RCL)
• Meter Hold (HOLD)
• Scale (°F/°C)
• Sensor Type
POWER OFF CONFIGURATION RETENTION:
Instrument retains last selected;
• Sensor Type
• °F/°C Scale
• 0.1°/1° Resolution
• Calibrator/Meter Mode
• Memory Contents (data storage and ramp parameters).
DIAGNOSTICS:
Display codes indicate following conditions;
• 'BAT': Low Battery
• 'OPEN': Open Input Thermocouple, RTD or Thermistor.
• 'E-1' (momentary): Invalid Keypad Entry
• 'E-2' (momentary): Ramp Function Not Installed
• 'HI': Temperature Reading or Keypad Entry Exceeds Rating
• 'LO': Temperature Reading or Keypad Entry Too Low For Accuracy
• LCD Test: During power-up, all segments and annunciators enabled momentarily
ENVIRONMENTAL LIMITS FOR OPERATING:
0°C to 50°C, less than 90% relative humidity
(R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 50°C.
ENVIRONMENTAL LIMITS FOR STORAGE:
-35°C to 60°C, less than 95% relative humidity
(R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 60°C.
POWER:
9 volt alkaline battery (NEDA 1604A).
BATTERY LIFE, CONTINUOUS:
10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable).
BATTERY INDICATOR:
Display indicates BAT when less than 10% of life remains.
DIMENSIONS, WEIGHT:
7.0" x 2.9 x 1.1". Net weight 12 oz.