Display reads E1 momentarily.
(1) This indicates that an invalid keypad
entry has been made. Review keystroke
sequence, or consult manual for input
instructions.
Display reads E2 momentarily.
(1) Attempt made to activate non-
installed ramp function.
NOTE:
RAMP Function enabled on Models 845 and 855 only.
Display reads HI.
(1) Meter-mode input-temperature
exceeds rating of selected sensor.
(2) Calibrator-mode keypad-entry
exceeds rating of sensor type.
Display reads LO.
(1) Meter-mode input-temperature
too low for accurate measurement.
(2) Calibrator-mode keypad-entry too
low for accurate simulation.