SERVICE INFORMATION
Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence that follows.
This will ensure that adjustments to be stored in EEPROM go to the
correct locations.
1. Perform meter-mode adjustments and verifications per Table 3 (use Table 5 for RTD-Thermistor
models).
2. Press OPR key (Flashing 'RTD' annunciator turns off).
3. Install J1 jumper.
4. Set unit to CALIB mode, RTD-100, °F.
5. Switch current source to OPERATE mode.
6. Perform calibrator-mode adjustments and verifications per Table 4 (use Table 6 for RTD-
Thermistor models).
7. Turn off unit. Remove tape holding battery.
8. Re-install back cover. Resistance-function calibration is complete.
Table 3: Calibration of Meter-Mode Resistance Functions (RTD-100/1000)
STEP
SENSOR
SELECT
UNIT
OUTPUT
ADJUST
VERIFY
DMM
READING*
1
RTD-100
390.48Ω
P1**
1562.0±0.5
°
F
2
RTD-100
100.00Ω
√
32.0±0.1
°
F
3
RTD-100
18.52Ω
√
-328.0±0.2
°
F
4
RTD-1000
3904.8Ω
P1**
1562.0±0.5
°
F
5
RTD-1000
1000.0Ω
√
32.0±0.1
°
F
6
1,000Ω
900.0Ω
P1
900.00Ω
7
1,000Ω
1.00Ω
√
1.00±0.10Ω
8
10,000Ω
9000.0Ω
P1
9000.0Ω
9
10,000Ω
10.0Ω
√
10.0±1.0Ω
*
exclusive of noise
** adjust as necessary to obtain steps 2,5.