6.7.2.1.2. Example RTD Calibration
Apply test inputs to meter
The calibration can now be tested. Apply an actual signal to the meter press “Test” to
check if the meter reads the correct value. Enter “Submit” to enter new factors if the
calibration was successful.
6.7.2.2. Simulate Flow
(Path: Diagnostics > Commissioning > Simulate Flow)
This option allows the meter to run in a simulated ("forced") state. While in this state,
the values displayed
are not
real measured information. To simulate:
(1) Enter a desired flow velocity; typical 700-1700
(2) Enter a desired velocity of sound (VOS); typical 0-20
(3) Select a desired flow profile from the drop down menu (0-8)
(4) Press button to start simulation
A "Forced Values" indicator will appear (if not already present) in the bottom right corner
of the display denoting simulation started.
Profile
Behavior
0
Standard ("Normal") operation
1 - 8
Various non-standard flow
profile types