Issue/Rev. 0.1 (11/18) ║ MNKS025 • Page 117
Parameter
Name
Menu
Range
Nominal
Description
Meter Tag
Communications
Up to 16
characters
(factory set to electronics
serial number)
Unique tag name designated for this
meter; used for identification.
Max Velocity of
Sound
Limits
300 m/s
4000 m/s
Depends on meter type
Alarm when measured velocity of
sound exceeds this value.
Min Velocity of
Sound
Limits
300 m/s
4000 m/s
Depends on meter type
Alarm when measured velocity of
sound falls below this value.
Min Signals Used
Limits
1 %
99 %
50% typ.
Alarm when percentage of signals
used falls below this value.
Max RX Gain
Limits
-12 dB
48 dB
30 db typ.
Alarm when receiver gain exceeds
this value.
Max RX Gain Dif-
ference
Limits
0 dB
64 dB
10 dB typ.
Alarm when the gain on an individual
path differs more than this value
compared to the median gain.
Max VOS Devia-
tion
Limits
0 m/s
1000 m/S
Depends on meter type
Alarm when the velocity of sound on
an individual path differs more than
this value compared to the median
velocity of sound.
Min Signal to
Noise Ratio
Limits
0 dB
60 dB
20 dB typ.
Alarm when the used signal to noise
ratio falls below this value.
Max Turbulence
Level
Limits
0 %
50 %
20 % typ.
Alarm when the turbulence level
exceeds this value.
Max Swirl/Cross-
flow Deviation
Limits
0%
70 %
10 %
Alarm when the transversal flow
exceeds this value.
Max flow rate
Limits
5 m/s
40 m/s
Depends on meter type
Alarm when the flow velocity
exceeds this value.
Min flow rate
Limits
-40 m/s
-5 m/s
Depends on meter type
Alarm when the flow velocity falls
below this value.
Max Profile Flat-
ness Deviation
Limits
0 %
50 %
20 %
Alarm when the profile flatness
deviates this amount.
Max Profile Sym-
metry Devia-
tion
Limits
0 %
50 %
10 %
Alarm when the profile symmetry
deviates this amount.
Manual (force)
Values
Modes
-off
-on
-off
Controls whether manual values are
allowed to be used.
Serial Port Func
-
tion
Communications
-none
-modbus RTU
-modbus ASCII
-modbus RTU
Selects the serial port protocol.
Baud Rate Selec
-
tion
Communications
-1200
-2400
-4800
-9600
-19200
-38400
-57600
-115200
-9600
Select the serial port baud rate.
Parity Selection
Communications
-none
-odd
-even
-none
Select the serial port parity.
MPU Series C I/O/M Manual
Appendix A – Description of Legally Relevant Parameters
Summary of Contents for Smith Meter MPU 1600c
Page 10: ...Page 10 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 34: ...Page 34 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 38: ...Page 38 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 98: ...Page 98 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 106: ...Page 106 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 112: ...Page 112 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 119: ...Issue Rev 0 1 11 18 MNKS025 Page 119 Page intentionally left blank ...