Issue/Rev. 0.1 (11/18) ║ MNKS025 • Page 45
6.4.2.2. Network Settings
(Path: Settings > Network Settings)
Configure the necessary Ethernet settings for network communication.
• DHCP – Meter will automatically be assigned an IP per the network's DNCP
server.
• Fixed – allows for manual configuration of the meter's network settings.
• Saved button – accepts new configuration.
NOTE:
These changes will not take effect until the meter has been power cycled.
6.4.2.3. Flow Calibration
(Path: Settings > Flow Calibration)
The menu allows for implementation of a meter factor that is variable with the flow rate
over up to 16 meter factor nodes. Each node is comprised of 2 parameters: a flow rate
and a corresponding meter factor.
Calibration node 1 sets the meter factor for all flow rates at or below the entered flow
rate. Additional calibration nodes can be entered in order from low to high flow rate in
order to vary the meter factor with flow rate.
Not all 16 available calibration nodes need to be configured. The last configured cali-
bration node sets the meter factor for all flow rates at or above the entered flow rate. If
a single calibration node is configured then this sets the meter factor for all flow rates
above and below the flow rate set point.
Calibration Node Entry Guidelines:
1) All 16 of the calibration nodes do not need to be configured. Start the calibration
range at Node 1 and leave unused nodes set to (0,0).
2) Use the single calibration Node 1 to set a fixed meter factor valid for all flow
rates.
3) Program additional nodes starting at the lowest flow rate at Node 1 progressing
to the highest.
4) Program nodes contiguously (don’t leave unused nodes in the middle of the
list).
5) The first configured node in the list sets the meter factor for all flow rates at or
below the entered flow rate
MPU Series C I/O/M Manual
Web User Interface
Summary of Contents for Smith Meter MPU 1600c
Page 10: ...Page 10 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 34: ...Page 34 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 38: ...Page 38 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 98: ...Page 98 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 106: ...Page 106 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 112: ...Page 112 MNKS025 Issue Rev 0 1 11 18 Page intentionally left blank ...
Page 119: ...Issue Rev 0 1 11 18 MNKS025 Page 119 Page intentionally left blank ...