gently twisting and then pulling it. Check
that the soft cannula does not exceed the
introducer needle.
CAUTION:
Before positioning AutoSoft™ 90
over a clean insertion site, place the tubing
in the slot to ensure that the tubing is not
caught underneath the infusion set during
insertion.
12.
Insert AutoSoft™ 90:
Position
AutoSoft™ 90 over the insertion
site. Simultaneously press the round
indentations on both sides of the insertion
device to insert AutoSoft™ 90.
13. Push gently on the center of the inserter to
secure the adhesive tape on the skin.
14. Remove the serter/introducer needle by
grasping the center and pulling it gently
back. Gently massage the tape onto the
skin. Fill the empty space in the cannula
with the following amount of insulin:
U-100 INSULIN:
6 mm: 0.3 units (0.003 ml)
9 mm: 0.5 units (0.005 ml)
15. Push the lid back in place until you hear
a “click.”
DISCONNECT
A. AutoSoft™ 90 allows you to temporarily
disconnect your pump without having to
change the infusion set. Hold the adhesive
in place by placing a finger just in front of
the cannula housing, and gently squeeze
the sides of connector. Pull the connector
needle straight out from the cannula
housing.
B. Insert the disconnect cover into the
cannula housing until you hear a “click”.
RECONNECT
C. If necessary, prime the infusion set until
medication dispenses from the needle.
CAUTION:
When priming the infusion
set, hold AutoSoft™ 90 with the needle
pointing down.
D. Remove the disconnect cover from the
cannula housing by pressing the grips.
Place a finger on the adhesive tape just in
front of the cannula housing.
E. Push the connector needle straight in until
you hear a “click”.
Summary of Contents for AutoSoft 90
Page 57: ...57 GREEK AutoSoft 90 AutoSoft 90...
Page 69: ...69 AutoSoft 90 1 3 1 2...
Page 73: ...73...
Page 75: ...75 B C AutoSoft 90 D E...
Page 76: ...76...
Page 78: ...78 15 AutoSoft 90 A B C AutoSoft 90 D E...
Page 85: ...85 AutoSoft 90 AutoSoft 90...
Page 89: ...89 6 7 8 9 10 11 12 13 4 5...