Using via Digital Data Communication
99
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
: Sets output voltage
Function:
Sets the output voltage
Format:
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <NRf>
Type:
SET command, QUERY command
Parameter:
30VType: 0.00 to 31.50/ 60VType: 0.00 to 63.00/ 500VType: 0.0 to 525/ 1000VType: 0 to 1050
Response:
Same as parameters
Restrictions:
Alarm response is returned for normal SET commands in the following conditions
·
While operating as a series operation slave unit
·
While operating as a parallel operation slave unit
<Ex.>
VOLT 30
VOLT?
30.00
[SOURce]:VOLTage:PROTection[:LEVel]
: Sets OVP voltage
Function:
Sets the OVP voltage
Format:
[SOURce]:VOLTage:PROTection[:LEVel] <NRf>
Type:
SET command, QUERY command
Parameter:
30VType: 0.30 to 33.00/ 60VType: 0.60 to 66.00/ 500VType: 5.0 to 550.0/ 1000VType: 1 to 1100
Response:
Same as parameters
Restrictions:
Alarm response is returned for normal SET commands in the following conditions
·
While operating as a parallel operation slave unit
<Ex.>
VOLT:PROT 33
VOLT:PROT?
33.00
Summary of Contents for HX-S-G2 Series
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......