Appendix B
B–40
TA I
NSTRUMENTS
TGA 2950
with from 80 to only 5 for each Hi-Res
TM
mode
while covering the majority of materials of
interest.
The next step is how to proceed after the first
Hi-Res
TM
run is complete. If the first run was
made at resolution setting 3.0, try the second one
at 4.0. Generally, increasing resolution setting
by a whole number will increase the time to
complete the TGA scan by a factor of 2 to 5
times. Therefore, you must consider whether
you can afford the added run time as well as the
extra setup time. This balance between in-
creased run time and adequate resolution is
usually the determining factor on what resolu-
tion setting to live with.
The benefit to having more range and resolu-
tion to the resolution setting than seems to be
necessary, is that on a rare occasion a material
requires a very fine adjustment or an extreme
treatment. Dont forget that the maximum
heating rate is a factor in determining resolution
as well as the resolution setting.
Lower resolution settings allow materials which
already have well-separated transitions to be
analyzed at super high heating rates such as
200
°
C/minute with excellent resolution in a
fraction of the time required at a constant rate of
20
°
C/minute. Resolution settings above 5.0 are
useful when exact decomposition temperatures
are needed ,or when the decompositions are
explosive in nature, or when overlapped transi-
tions are extremely close but highly temperature
selective.
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...