Appendix B
B–52
TA I
NSTRUMENTS
TGA 2950
at the rate of 2 times per second until the seg-
ment terminates normally or the limit is reached.
If the limit is reached during the execution of a
segment, the remaining portion of the segment is
skipped. Method execution then continues with
the next segment in the method.
The conditions for determining whether limits
have been reached for each type of limit are as
follows:
% - If the condition operator is “<“ and the
sample weight percent is less than or equal
to the limit percent, or if the condition
operator is “>” and the weight percent is
greater than or equal to the limit percent,
then the limit is reached. Values greater
than 100% are permitted to accommodate
weight gains.
%/min -If the condition operator is “<“ and the
derivative of sample weight percent is less
than or equal to the limit percent/minute, or
if the condition operator is “>” and the
derivative of weight percent is greater than
or equal to the limit percent/minute, then
the limit is reached. Negative values
indicateweight gain and positive values
indicate weight loss.
Abort segments are very versatile because they
can be used in front of any method segment
(including other abort segments) to dynamically
change the execution of a method. For example,
an abort segment can be used to control the
switching of a purge gas or activation of an
event relay based on the rate of reaction or the
amount of weight loss. An abort segment in
front of a repeat segment provides a mechanism
to terminate a loop prematurely. Abort segments
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...