Running Experiments
TA I
NSTRUMENTS
TGA 2950
3–18
It is important to maintain the proper ratio of
flow rates between the balance chamber and the
furnace housing. Having the separate balance
chamber purge prevents decomposition gases
from entering the balance chamber. The recom-
mended setting for the purge rate is 100 mL per
minute or less. When you use the standard TGA
2950 furnace, the flow distribution should be 40
percent to the balance chamber and 60 percent to
the furnace. When you use the TGA 2950 EGA
furnace, the flow distribution should be 10
percent to the balance chamber and 90 percent to
the furnace.
To maintain this flow distribution, you will need
to connect a flowmeter to each of the purge
fittings on the back of the TGA instrument. Set
the purge gas flow rates by adjusting these
meters. The PURGE port goes to the TGA
furnace, and the
BALANCE PURGE goes to the
balance chamber.
Before you start the purge gas, make sure that
the desired gas is connected to the purge ports,
that all lines are clear, and that your supply of
purge gas is sufficient for the experiment.
Always maintain constant purge flow rates and
distribution throughout your experiment;
changing the purge during an experiment can
affect the data.
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...