Appendix D
TA I
NSTRUMENTS
TGA 2950
D–18
Connecting
the Spectrometer
The TGA 2950 EGA furnace allows you to
connect a spectrometer, such as a FTIR spec-
trometer, to the instrument. To connect any
spectrometer, you will need to use a transfer line
(supplied by the spectrometer manufacturer) to
transport the gas evolved from the sample on the
TGA to the spectrometer.
•
The transfer line should be 1/8 inch in
diameter to connect with a 1/8-inch
Swagelok
TM
fitting on the exhaust gas
connection.
•
The transfer line should be made of heat-
resistant alloy capable of resisting
corrosion by the evolved gas and oxidation
at temperatures up to 1000°C.
•
The transfer line must pass through the
exhaust gas fitting and a glass branch tube
in the sample tube. It should end at a point
just short of the inside diameter of the
sample tube to ensure that the evolved gases
do not condense before entering the transfer
line.
•
The transfer line must be long enough to
allow flexible movement. It must
accommodate movement of the EGA
furnace up and down 3 5/8 inches to open
and close for sample loading and unloading.
(If the transfer line is not long enough, it
must be disconnected and reconnected each
time the furnace is opened and closed.)
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...