Technical Reference
TA I
NSTRUMENTS
TGA 2950
4–10
The stainless steel furnace housing has a purge
opening on either side to allow gas flow or
evacuation. Purge gas enters through a purge
fitting on the right side of the furnace housing. It
then passes through an opening in the heater,
flows around the sample pan, and exits through
openings on the opposite side of the furnace
housing and heater. From here, the evolved gas
can be sent through appropriate connections to
an external analyzer, if desired.
The heater is a resistance-wound unit of low
thermal mass alumina material that can be
heated and cooled rapidly. Controlled heating
rates of up to 100°C per minute can be obtained,
to an upper limit of 1000°C.
Cooling air enters through the holes in the base
of the furnace assembly at the completion of test
runs, if desired.
A Platinel II* thermocouple extends through
the bottom of the balance chamber, down along
the hang-down wire, and is positioned just above
the sample pan, where it monitors the sample
environment temperature.
The furnace base moves the furnace assembly
up around the sample pan to the closed position,
or down away from the sample pan to the open
position.
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...