
Appendix B
B–54
TA I
NSTRUMENTS
TGA 2950
again initiated, and so on until the final tempera-
ture is reached.
The stepwise heating technique can be easily
implemented by placing a ramp segment fol-
lowed by an isothermal segment into a “repeat to
final temperature” loop, and preceding the ramp
and isothermal segments with abort segments.
The abort segment preceding the ramp is setup
to terminate the ramp when the “%/minute” is
greater than a specified limit. The abort seg-
ment preceding the isothermal segment termi-
nates the holding period when the “%/minute” is
less than a second limit. An example of this
type of method is shown below.
1: Abort next seg if %/min > 0.5
2: Ramp 10
°
C/min to 700
°
C
3: Abort next seg if %/min < 0.05
4: Isothermal for 500 minutes
5: Repeat segment 1 til 700
°
C
The “%/min” limit for the isothermal segment is
selected to be equal to the baseline rate of
weight change encountered during the onset of
the transition of interest via a normal constant
heating rate scan of this material. (Be sure to
use the same ramp rate as that selected for
stepwise heating.) The “%/min” limit for the
ramp segment is selected to be about an order of
magnitude greater than that selected for the
isothermal segment (but not more than the
maximum rate of weight change encountered
during the transition of interest). The ramp final
temperature and the repeat final temperature are
set to the final experiment temperature. The
isothermal time is set to an arbitrary time, which
is sufficiently large such that the segment will
not terminate until the percent/minute limit has
been reached.
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...