TA I
NSTRUMENTS
TGA 2950
5–7
2. Unscrew and remove the water reservoir
cap.
3. Drain the coolant and flush out the system
as follows:
a.
Lift the heat exchanger and dump out
the contents of the water reservoir
bottle.
b.
Fill the bottle to 2/3 full with distilled
water only and replace the cap.
c.
Reconnect the heat exchanger cable and
water lines to the instrument cabinet.
d.
Turn on the
POWER
switch.
e.
Turn on the pump by activating Air
Cool and allow the water to circulate for
several minutes.
f.
Turn off the pump by deactivating Air
Cool, and check the clarity of the water
in the reservoir bottle.
g.
If the water clarity is still unacceptable,
disconnect the heat exchanger cable and
water lines from the instrument cabinet,
and repeat steps a through f.
h.
Continue repeating this procedure until
you are satisfied with the clarity of the
water in the bottle after it has circulated.
4. Dispose of the water and fill the bottle
with TGA Conditioner (PN 952377.901)
and fresh distilled water as directed in
Chapter 2.
Routine Maintenance
Summary of Contents for TGA 2950
Page 20: ...TA INSTRUMENTS TGA 2950 xx ...
Page 22: ...Introducing the TGA 2950 TA INSTRUMENTS TGA 2950 1 2 ...
Page 36: ...TA INSTRUMENTS TGA 2950 Installing the 2950 2 2 ...
Page 90: ...Running Experiments TA INSTRUMENTS TGA 2950 3 24 ...
Page 92: ...Technical Reference TA INSTRUMENTS TGA 2950 4 2 ...
Page 112: ...Technical Reference TA INSTRUMENTS TGA 2950 4 22 ...
Page 136: ...TA INSTRUMENTS TGA 2950 Maintenance and Diagnostics 5 24 ...
Page 142: ...Appendix B B 4 TA INSTRUMENTS TGA 2950 ...
Page 152: ...Appendix B B 14 TA INSTRUMENTS TGA 2950 Figure B 2 Figure B 3 ...
Page 153: ...TA INSTRUMENTS TGA 2950 B 15 Hi ResTM Option Figure B 4 Figure B 5 ...
Page 169: ...TA INSTRUMENTS TGA 2950 B 31 Hi ResTM Option Figure B 7 Figure B 8 ...
Page 188: ...Appendix B B 50 TA INSTRUMENTS TGA 2950 Figure B 11 Sensitivity Setting Too High ...
Page 205: ...TA INSTRUMENTS TGA 2950 B 67 Hi ResTM Option Figure B 16 Figure B 17 ...
Page 211: ...TA INSTRUMENTS TGA 2950 B 73 Hi ResTM Option Figure B 22 Figure B 23 ...
Page 212: ...Appendix B B 74 TA INSTRUMENTS TGA 2950 Figure B 24 Figure B 25 ...
Page 214: ...Appendix B B 76 TA INSTRUMENTS TGA 2950 ...
Page 216: ...Appendix C TA INSTRUMENTS TGA 2950 C 2 ...
Page 218: ...Appendix C TA INSTRUMENTS TGA 2950 C 4 Figure C 1 Auto TGA ...
Page 236: ...Appendix D TA INSTRUMENTS TGA 2950 D 2 ...
Page 258: ...Appendix D TA INSTRUMENTS TGA 2950 D 24 ...
Page 264: ...Index TAINSTRUMENTSTGA2950 I 6 ...