![Swann AMI Codes-II Operator'S Manual Download Page 49](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948049.webp)
AMI Codes-II
Maintenance
A-96.250.571 / 050517
47
6
Put the canister [E] into the holder [G].
7
Remove the screw cover, insert the suction lance [A] and tight-
en the screw cover.
Prepare
Oxycon
On-line Buffer
1
Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2
Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3
Slowly pour the content of one bag of buffer Oxycon On-line
(240 gr) into the canister.
Avoid splashing!
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well.
6
Put canister [F] into holder [G].
7
Remove the screw cover and insert the suction lance [B] and
tighten the screw cover.
8
Fill reagent system. See