AMI Codes-II
Troubleshooting
E011
Temp. shorted
– check wiring of temperature sensor, see
– check temperature sensor
E012
Temp. disconnected
– check wiring of temperature sensor, see
– check temperature sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value in menu
E014
Case Temp. low
– check case/environment temperature
– check programmed value in menu
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
E018
Reagent Pump
– shut off power
– check wiring, see
E019
Photometer not con-
nected
– shut off power
– check wiring of photometer, see
E020
Photometer dirty
– check process,
– clean photometer, see
E021
DIS. invalid
– This error appears after start-up and will
disappear after the first valid
measurement is finished.
E022
Reagent empty
– refill reagents, see
E023
Cleaning solution
– refill cleaning solution
E024
Input active
– See If Fault Yes is programmed in Menu
E026
IC LM75
– call service
E028
Signal output open
– check wiring on signal outputs 1 and 2
Error
Description
Corrective action