AMI Codes-II
Program List and Explanations
A-96.250.571 / 050517
87
5.3.1.2.1
Alarm High:
If the measured value rises above the alarm high
value, the alarm relay is activated and E003 is displayed in the
message list.
Range: 0–14.00 pH
5.3.1.2.21
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list
Range: 0–14.00 pH
5.3.1.2.31
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value
Range: 0–14.00 pH
5.3.1.2.41
Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 sec
5.3.1.3
Sample Flow:
Define at which sample flow a flow alarm should be
issued.
5.3.1.3.1
Flow Alarm:
Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
NOTICE:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.3.x
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 100–600 B/s
5.3.1.3.x
Alarm Low:
If the measuring values falls below the programmed
value E010 will be issued.
Range: 5–80 B/s