AMI Codes-II
Product Description
A-96.250.571 / 050517
13
2.1.
Instrument Specification
Power Supply
Voltage:
Power consumption:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
min. 10 l/h
5–50 °C (41–122 °F)
0.15–2 bar (2–28 PSI)
pressure free
NOTICE:
No oil, no grease, no sand.
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Tube 6 x 8 mm
1/2” hose nozzle for flexible tube
diam. 20x15 mm
Measuring
Range
Ozone
0.000–1.000 ppm
Accuracy
± 0.005 ppm
HOCL, free chlorine, monochloramine
0.00–1.00 ppm
1.00–3.00 ppm
3.00–5.00 ppm
± 0.01 ppm
± 0.06 ppm
± 0.2 ppm
Chlorine dioxide, iodine, bromine
0.00–2.00 ppm
2.00–6.00 ppm
± 0.02 ppm
± 0.12 ppm