AMI Codes-II
Troubleshooting
A-96.250.571 / 050517
67
Error
Description
Corrective action
E001
DIS. Alarm high
– check process
– check programmed value in menu
E002
DIS. Alarm Low
– check process
– check programmed value in menu
E003
pH Alarm high
– check process
– correct/calibrate pH sensor, see
– check programmed value in menu, see
E004
pH Alarm low
– check process
– correct/calibrate pH sensor, see
– check programmed value in menu, see
E005
DIS. too high
– check process
E007
Sample Temp. high
– check sample temperature
– check programmed value in menu
E008
Sample Temp. low
– check sample temperature
– check programmed value in menu
E009
Sample Flow high
– check sample input pressure
– readjust sample flow
– check programmed value in menu
E010
Sample Flow low
– check sample input pressure
– readjust sample flow
– clean instrument, see
– check programmed value in menu