![Swann AMI Codes-II Operator'S Manual Download Page 53](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948053.webp)
AMI Codes-II
Maintenance
A-96.250.571 / 050517
51
Process pH
Use a Chematest 25 photometer (or equivalent) to determine the
sample pH value.
NOTICE:
Make sure your reference instrument is calibrated
correctly!
Possible error message see
.
Enter the correct value
with the [
] or [
]
key.
3.1
Calibration
Service
Maintenance
Set Time 01.01.05 16:30:00
Calibration
Simulation
Cleaning
Enter
3.1.2
Calibration
Process DIS.
Process pH
Standard pH
Process pH
Offset
x mV
Save
<Enter>
Process Value
7.78 pH
3.1.2.4
Current Value
7.78 pH
Process pH
Current Value
7.78 pH
Offset
x mV
Save
<Enter>
Process Value
7.70 pH
3.1.2.4
3.1.2.5
Process pH
Current Value
7.70 pH
Offset
y mV
Slope
x.xx mV
Calibration successful
Enter
3.1.2.5
Process pH
Current Value
7.70 pH
Offset
y mV
Process Value
7.70 pH
Save
<Enter>