12
A-96.250.571 / 050517
AMI Codes-II
Product Description
Time
interval of a
measurement
The measuring interval can be set between 1 and 12 minutes. The
time sequence of a measurement with a measuring interval of
5 min is shown in the diagram below.
The blue bar represents the sample which flows continuously
through the photometer. A short time before the measurement
starts, a zero point measurement is performed. Then the peristaltic
pump starts and a small portion of the reagents [J] and [K] is
pumped into the mixing chamber. Shortly after, if the mixture is in
the photometer, the sample is measured.
J
K
OXYCON ON-LINE DPD
OXYCON ON-LINE Buffer Solution
Zero point measurement
Sample measurement
1
0
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19 20
m
J
K