SVAN 979 User Manual
Selecting
Lin
is required for obtaining the true RMS value of the measured signal according to IEC 61672-
1:2013. When this option is selected the value of the
Leq
,
SEL
,
LEPd
and
L(den)
results do not depend on
the detector time constant:
Fast
,
Slow
or
Impulse
(the results are displayed without the indicator of the
detectors selected in the profiles). In this case, the indicator
Linear
(or
L
) is displayed during the result
presentation.
Selecting
Exp
enables fulfilling the requirements of other standards for time averaged
Leq
measurements.
When this option is selected the value of the
Leq
,
SEL
,
LEPd
and
L(den)
results depend on the detector time
constant (
Slow
,
Fast
,
Impulse
). Results are displayed with the indicator of the detector type selected in the
profiles (
path: <Menu> / Measurement / Profiles
).
Detector for statistics
The
Statistics
parameter defines the method for calculation of statistical results
Ln
. Statistics for profiles will be calculated based on RMS results with linear
detector (
Lin
) or exponential detector (
Exp
), e.g.
Impulse
,
Fast
or
Slow
,
defined in the
Profiles
list, regardless of how main results are measured. Default
value:
Lin
.
Day time limits
The
Day Time Limits
parameter defines the day and night time limits required
by the local standards. These limits are used for the calculation of the
Lden
function (see Appendix D for definition). Two options are available:
6h-18h
and
7h-19h
. By default, it is set to
6h-18h
.
The
Rolling Time(x)
parameter defines „time window” for the "Rolling Leq"
calculation. The Rolling Leq are presented as LR+<time window>. For example,
if the
Rolling Time
is equal to 30 minutes, the appropriate result will be named
as
LR30
and will be calculated each second as Leq of last 30 minutes. If the
Rolling Time
is in seconds, for example is equal to 50s, the result name will
have letter "s" (
LR50s
), to make it different from 50 minutes LR result (
LR50
).
Note: In case of vibration measurements, positions Statistics, Day Time Limits and
Rolling Time(x) are not presented in the General Settings list
.
You
can
easily
access
the
General Settings
screen
during
the
measurement performance from the result
view. It is necessary to enter some result
field (for example,
RMS
) with the ▲ / ▼ or
◄ / ► key and press
<ENTER>
.
<ENT>
4.2 S
ETTING MEASUREMENT TRIGGER
–
M
EASUREMENT
T
RIGGER
The
Measurement Trigger
list allows you
to set up parameters of the measurement
trigger, that switches the measurement
process in case the trigger condition is met.
The
Measurement Trigger
can be switched
Off
or On by selecting its type in the
Trigger
position.
<ENT>
Trigger is switched on if one of its six available modes is selected in the
Trigger
position:
Slope +
,
Slope
–
,
Level +
,
Level
–
,
Grad +
or
External
. When the trigger is selected, additional parameters can be defined: the
measurement result that is checked for a trigger condition (
Source
), its threshold value (
Level
) and the rate of
its changing (
Gradient
).
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...