Antialiasing filter
Built-in antialiasing filter (eighth-order elliptic type) ensuring correct sampling of the measured signal.
Pass band(-1 dB)
22.200 kHz or 44.4 kHz
Pass band(-3 dB)
23.520 kHz or 47.0 kHz
Stop band
26.256 kHz or 52.53 kHz
Attenuation in the stop band
> 100 dB.
Reference conditions
•
Reference frequency
79.58 Hz
•
Reference temperature
+23°C
•
Reference relative humidity
50%
•
Reference range
HIGH.
Pre-heating time
1 minute (for 0.1 dB accuracy).
Typical stabilization time after change in environmental conditions is 1 minute.
Note: When the instruments are moved from a warm environment with high humidity,
to a colder environment, care should be taken not to produce condensation inside
the instruments. In this case, much longer stabilization periods may be necessary.
Digital filters
High-pass
filters
HP
filter
(see Chapter C.5 for the filter characteristics)
HP1
filter
(see Chapter C.5 for the filter characteristics)
HP3
filter
(see Chapter C.5 for the filter characteristics)
HP10
filter
(see Chapter C.5 for the filter characteristics)
Frequency weighting
filters (filter includes Band Limiting filter).
(Conforms to ISO 8041-1:2017, see Chapter C.5 for the frequency response characteristics)
•
Wh
from 0.8 Hz to 4000 Hz
Integrating filters
(see Chapter C.5 for the frequency response characteristics):
•
Vel1
from 0.2 Hz to 4100 Hz
•
Vel3
from 0.2 Hz to 4100 Hz
•
Vel10
from 0.2 Hz to 4100 Hz
•
Dil1
from 0.1 Hz to 260 Hz
•
Dil3
from 0.2 Hz to 510 Hz
•
Dil10
from 1 Hz to 2050 Hz
Special filter
Filter for the evaluation of the machinery condition:
•
VelMF
from 0.2 Hz to 4100 Hz; conforms to the ISO 10816 standard (see Chapter C.5 for the frequency
response characteristics)
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...