![Svantek SVAN 979 User Manual Download Page 186](http://html.mh-extra.com/html/svantek/svan-979/svan-979_user-manual_1409285186.webp)
186
SVAN 979 User Manual - Appendixes
A.2.4. Examples
The exemplary averaged results of the instrument’s response after sending to it the following sequence
of characters:
#2,a,1;
coming from the first profile are given below:
a) for the case of the
SLM
mode:
#2,a,1,v1,V0,T39,P125.4,M107.0,N20.6,S81.7,R102.1,U118.0,B(4)112.1,I(480)102.1,Y103.9,Z105.4,
L(01)107.9,L(10)107.6,L(20)107.2,L(30)102.8,L(40)99.0,L(50)96.7,L(60)82.5,L(70)54.5,L(80)20.9,
L(90)20.4,g90.8,h94.4,C1,F2017-11-24,G16:26:03,O?,K?;
b) and for the case of the
VLM
mode:
#2,a,1,v0,V0,T1,P93.9,Q99.7,M45.6,R45.6,H85.0,C1,F2017-11-24,G17:00:01;
Note: The presented above order of the measurement results sent out by the instrument does not
depend on the sequence of the characters sent to the unit.
Example:
After sending to the instrument the string:
#2,i,1,T?,R?,V?,P?,L?;
the unit sends out the instantaneous results of measurement coming from the first profile in predefined,
described above, order:
#2,i,1,V0,T39,P125.4,R102.1,L(01)107.9,L(10)107.6,L(20)107.2,L(30)102.8,L(40)99.0,L(50)96.7,
L(60)82.5,L(70)54.5,L(80)20.9,L(90)20.4;
Note: All bytes of that transmission are ASCII characters.
17.9 A.3.
F
UNCTION
#3
–
R
EAD OUT OF THE MEASUREMENT RESULTS IN
1/
X
OCTAVE
OR
FFT
MODE
#3
function enables one to read out the current measurement results in
1/x OCTAVE
or
FFT
mode.
#3 function
format is defined as follows:
#3,A; -
averaged spectrum
#3,I; -
instantaneous spectrum
#3,M; -
max spectrum
#3,N; -
min spectrum
The device responds, sending the last measured spectrum (when the instrument is in STOP state) or currently
measured spectrum (when the instrument is in RUN state) in the following format:
#3;
<Status Byte>
<LSB of the transmission counter> <MSB of the transmission counter>
<LSB data byte> <MSB data byte>
(...)
<LSB data byte> <MSB data byte>
Status Byte
gives the information about the current state of the instrument.
b7
b6
b5
b4
b3
b2
b1
b0
where:
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...