Measurement
Trigger
Screen that enables configuring triggering of the
measurement/integration process with parameters:
Trigger, Source, Level
and
Gradient
.
Chapter
Trigger
Position that switches
Off
or on the measurement trigger
by selecting its type:
Slope +, Slope
–, Level +, Level –,
Grad +
or
External
.
If the instrument is waiting for the trigger condition, the
appropriate trigger icon is flashing on the display
alternatively with the „measurement” icon.
Chapter
Slope +
Type of trigger that starts the measurement/integration
by the duration of the
Integration Period
on condition:
rising value of the RMS result (
Source
) integrated during
0,5 ms passes above the threshold value (Level).
Chapter
Slope -
Type of trigger that starts the measurement/integration
by the duration of the
Integration Period
on condition:
falling value of the RMS result (
Source
) integrated during
0,5 ms passes below the threshold value (Level).
Chapter
Level +
Type of trigger that starts the 1-second measurement/
integration on condition: value of the RMS result (
Source
)
integrated during 0,5 ms is greater than the threshold
value (
Level
). In other cases, the instrument continues
checking the trigger condition every 0,5 mc. During one
measurement cycle the instrument performs as many 1-
second
integrations
as
many
seconds
the
Integration Period
consists and stops the measurement
cycle.
Chapter
Level -
Type of trigger that starts the 1-second measurement/
integration on condition: value of the RMS result (
Source
)
integrated during 0,5 ms is lower than the threshold value
(
Level
). In other cases, the instrument continues
checking the trigger condition every 0,5 mc. During one
measurement cycle the instrument performs as many 1-
second
integrations
as
many
seconds
the
Integration Period
consists and stops the measurement
cycle.
Chapter
Grad +
Type of trigger that starts the 1-second measurement/
integration on condition: value of the RMS result (
Source
)
integrated during 0,5 ms is greater than the threshold
value (Level) and the gradient of this Source is greater
than the threshold level (
Gradient
). In other cases, the
instrument continues checking the trigger condition every
0,5 mc. During one measurement cycle the instrument
performs as many 1-second integrations as many
seconds the
Integration Period
consists and stops the
measurement cycle.
This type of trigger has the same logic as
Level +
trigger,
but the trigger condition requires also gradient level to be
exceeded.
Chapter
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...