Every time you start a new measurement
with the
<Start>
key, a new file will be
created and saved in the point directory
after the measurement is completed.
If you break the measurement with the
<Stop>
key, the measurement file will not
be created.
<Start>
If you delete the measurement in the
Point
view, the appropriate file will be deleted
from the point directory.
=>
Note: Project, Area and Source directories can be created, renamed or deleted via the File Manager
- see Chapter
Note: You cannot delete or rename point directories and measurement file via the File Manager.
You can delete point directory and measurement file using the mechanism described in
Chapter
14.4.3
Averaging results
STIPA measurements require averaging of STI/CIS indexes. This averaging concerns two aspects:
1) averaging of results for a measurement
point.
Averaged
STI
/
CIS
for the measurement point
is displayed in the left upper corner of the
Point
view and in the measurement points list
of the
Area
view.
2) averaging of results for a tested area.
Averaged
STI
/
CIS
for the tested area is
displayed in the left upper corner of the
Area
view and in the second line of the
Area
summary
view.
According to most standards, it is good practice to average the results for measurement points for two or three
measurements. In case of noise fluctuations during measurements, it is recommended to make not less than
three measurements and check whether the spread of the STI index does not exceed 0.03.
14.4.4
STIPA settings
STIPA measurements are configured in the
STIPA Settings
and
Ambient noise
screens. To configure a
STIPA project follow these steps:
1. Select the
STIPA Settings
position in
the
Measurement
section and open it
with the
<ENTER>
key.
<ENT>
Summary of Contents for SVAN 979
Page 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Page 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Page 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Page 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Page 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Page 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Page 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Page 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Page 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Page 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Page 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...