SVAN 948 USER MANUAL
4 - 23
Selection of measurement parameters - MEASURE SETUP
All settings made in
MEASURE SETUP
are common for sound and vibration level meters. The
MEASURE SETUP
sub-list is opened after the selection of the
MEASURE SETUP
text from the
INPUT
list by means of the
<
>
,
<
>
(or
<
>
,
<
>
) push-buttons and pressing the
<ENTER>
one.
The
MEASURE SETUP
sub-list consists of the parameters, which can be set, switched on or off, namely:
the delay of the start of measurements (
START DELAY
), the integration time (
INT. TIME
), the repetition
of the measurement cycles (
REP. CYCLE
) and the buffer’s step (
BUF. STEP
). In order to change the
displayed inversely parameter the user has to press the
<
>
,
<
>
push-buttons. The confirmation of any
change made in the sub-list requires pressing the
<ENTER>
push-button, which simultaneously closes
the sub-list. The
MEASURE SETUP
sub-list is closed ignoring any changes made, after pressing in any
time the
<ESC>
push-button.
The view of the displays with the MEASURE SETUP sub-list opened
(path: MENU / INPUT / MEASURE SETUP)
Setting time delay before the start of measurements - START DELAY
The
START DELAY
position defines the delay from the pressing the
<START / STOP>
push-button
to the start of the measurements (the digital filters of the instrument analyse constantly the input signal
even when the measurements are stopped). This delay can be set from
1 second
to
60 seconds
(with
1 second step by means of the
<
>
,
<
>
push-buttons and with 10 seconds step with the
<
>
,
<
>
push-buttons pressed together with the
<SHIFT>
one). The
<ENTER>
push-button must be pressed for
the confirmation of the selection, which closes simultaneously the
MEASURE SETUP
sub-list.
The view of the displays with the MEASURE SETUP sub-list opened
(path: MENU / INPUT / MEASURE SETUP);
the setting of the START DELAY
Notice: The minimum delay time is equal to 1 second. It protects from the registration
by the instrument the effects of the push-button pressing.
In the
CALIBRATION
mode the delay time
is equal to 5 seconds.
Setting the integration time - INT. TIME
The
INT. TIME
position defines the time period in which the signal is being averaged during the
sound level measurements. The required parameter can be set by means of the
<
>
,
<
>
and
confirmed by the
<ENTER>
push-button.
The integration time (
INT.TIME
) can be set (by pressing the
<
>
or
<
>
with
<SHIFT>
push-
buttons):
Summary of Contents for SVAN 948
Page 1: ...SVAN 948 SOUND AND VIBRATION DATA LOGGER USER S MANUAL SVANTEK Sp z o o WARSAW June 2004...
Page 19: ...SVAN 948 USER MANUAL 2 6 The view of the rear panel of the SVAN 948 instrument in 1 1 scale...
Page 25: ...SVAN 948 USER MANUAL 3 6 Control diagram of the INPUT list...
Page 28: ...SVAN 948 USER MANUAL 3 9 Control diagram of the DISPLAY list...
Page 30: ...SVAN 948 USER MANUAL 3 11 Control diagram of the FILE list...
Page 32: ...SVAN 948 USER MANUAL 3 13 Control diagram of the AUX FUNCTION list...
Page 36: ...SVAN 948 USER MANUAL 3 17...
Page 37: ...SVAN 948 USER MANUAL 3 18 Control diagram of the SETUP list...