SVAN 948 USER MANUAL
_
5 - 56
Operations in buffer
The buffer is the selected part of the instrument’s memory dedicated for storing the huge number of
the measurement results. Its capacity is ca 3.4 Mbytes (for the unit with 8 MB of internal memory). The
buffer operation is strongly dependent on the operation mode of the instrument (the selected function).
Notice:
The new file in the buffer is created after each start of the new measurement. The
contents of the buffer (all registered files) are erased after execution the
CLEAR BUFFER
operation from
the
FILE
list. It is not possible to erase only one or few selected files from the buffer!
1. In the
Sound Level Meter
mode selected results from each of four channels can be logged in the
buffer’s file with the time step defined in the
BUF. STEP
position. If the value set in this position is
equal to 1 sec, thus for three profile results logging the buffer is fulfilled after 168 hours (or 7 days) of
the continuous work. If the results from only one profile are saved the buffer will be filled after
500 hours (21 days) of work!
EXAMPLE 1.
Assuming that the following settings are selected:
MEASURE SETUP
sub-list
INT. TIME: 10 m
,
REP. CYCLE: 1
,
BUF. STEP: 1 s
,
BUFFERS SETUP
sub-list
CHANNEL 1: BUFFERS: PEAK
,
CHANNEL 2: BUFFERS: RMS
,
CHANNEL 3: BUFFERS: None
.
CHANNEL 4: BUFFERS: None
Then after 10 minutes (the value set in
INT. TIME
position) and after one measurement cycle (the
value set in the
REP. CYCLE
position) the current file in the buffer will contain 600
PEAK
values from the
channel 1 and 600 one second
RMS
values (one second set in the
BUF. STEP
position) from the
channel 2.
EXAMPLE 2.
Assuming that the instrument set-up is as follows:
MEASURE SETUP
sub-list
INT. TIME: 1 m
,
REP. CYCLE: 1
,
BUF. STEP: 7 s
,
BUFFERS SETUP
sub-list
CHANNEL 1: BUFFERS: PEAK
,
CHANNEL 2: BUFFERS: RMS
,
CHANNEL 3: BUFFERS: None
.
CHANNEL 4: BUFFERS: None
.
Then after the end of the measurement the current file in the buffer will contain 8
PEAK
values (for
each 7-seconds period set in the
BUF. STEP
position) from the channel 1 and 8
RMS
values from the
channel 2.
Notice:
The setting in a profile
BUFFER: ON
will result in bypassing the registration of the
measurement results from this profile in the current file in the buffer. In the same time it will create more
space in the buffer for the results from other profiles (independent instruments)!
2.
In
1/1 OCTAVE
or
1/3 OCTAVE
analysis one selected result from each of three profiles together with
the spectra can be logged in the current file of the buffer with the step selected in the
BUF. STEP
position.
Summary of Contents for SVAN 948
Page 1: ...SVAN 948 SOUND AND VIBRATION DATA LOGGER USER S MANUAL SVANTEK Sp z o o WARSAW June 2004...
Page 19: ...SVAN 948 USER MANUAL 2 6 The view of the rear panel of the SVAN 948 instrument in 1 1 scale...
Page 25: ...SVAN 948 USER MANUAL 3 6 Control diagram of the INPUT list...
Page 28: ...SVAN 948 USER MANUAL 3 9 Control diagram of the DISPLAY list...
Page 30: ...SVAN 948 USER MANUAL 3 11 Control diagram of the FILE list...
Page 32: ...SVAN 948 USER MANUAL 3 13 Control diagram of the AUX FUNCTION list...
Page 36: ...SVAN 948 USER MANUAL 3 17...
Page 37: ...SVAN 948 USER MANUAL 3 18 Control diagram of the SETUP list...