SVAN 948 USER MANUAL
4 - 3
4.1. CALIBRATION
The instrument is factory calibrated with the supplied accelerometer and microphone for standard
environmental conditions. Each of four channels should be calibrated separately. For situations where the
absolute vibration or sound level value is important, a calibration of the measurement channel should be
conducted. This is due to the fact that the accelerometer or microphone sensitivity is a function of
temperature, ambient pressure and humidity.
In order to select a calibration function the user has to enter the
FUNCTION
list by pressing the
<MENU>
push-button, selecting the
FUNCTION
text by means of the
<
>
,
<
>
(or
<
>
,
<
>
) push-
buttons and pressing the
<ENTER>
. The
FUNCTION
list contains two sub-lists:
MEASUR. FUNCTION
and
CALIBRATION
. The user has to open the
CALIBRATION
sub-list (to select the
CALIBRATION
text
using the
<
>
,
<
>
or
<
>
,
<
>
push-buttons and press the
<ENTER>
one, when this text is displayed
inversely). Then the user has to choose channel to calibration.
a)
b)
c)
d)
The view of the displays with the main list, the FUNCTION text selected (a), the
FUNCTION
list opened (b),
the CALIBRATION text selected (c) and CHANNEL selected (d)
In this sub-list two options are available: the calibration of the measurement channel by the
introduction of the accelerometer’s or microphone’s sensitivity from theirs calibration cards
(
BY SENSITIVITY
) or by performing the calibration measurement (
BY MEASUREMENT
).
The view of the displays in the CALIBRATION
(path: MENU / FUNCTION / CALIBRATION / CHANNEL X
)
sub-list
with the existing possibilities selected
Note:
The calibration level and the calibration result is expressed in different units depending
on the settings of the instrument. The metric or non-metric vibration units are set in the
VIBRATION UNITS
window of the
SETUP
list. Additionally, the linear or logarithmic units are set in the
DISPLAY SCALE
sub-list of the
DISPLAY
list.
Summary of Contents for SVAN 948
Page 1: ...SVAN 948 SOUND AND VIBRATION DATA LOGGER USER S MANUAL SVANTEK Sp z o o WARSAW June 2004...
Page 19: ...SVAN 948 USER MANUAL 2 6 The view of the rear panel of the SVAN 948 instrument in 1 1 scale...
Page 25: ...SVAN 948 USER MANUAL 3 6 Control diagram of the INPUT list...
Page 28: ...SVAN 948 USER MANUAL 3 9 Control diagram of the DISPLAY list...
Page 30: ...SVAN 948 USER MANUAL 3 11 Control diagram of the FILE list...
Page 32: ...SVAN 948 USER MANUAL 3 13 Control diagram of the AUX FUNCTION list...
Page 36: ...SVAN 948 USER MANUAL 3 17...
Page 37: ...SVAN 948 USER MANUAL 3 18 Control diagram of the SETUP list...