Selection of detector’s type in the LEQ (RMS) calculations - RMS INTEGRATION
The
RMS INTEGRATION
(path: MENU / SETUP / RMS INTEGRATION)
position enables the user to
select the detector type for the calculations of the
LEQ
function (in the case of sound measurements) or
the
RMS
function (in the case of vibration measurements). In order to enter this position the user has to
select the
RMS INTEGRATION
text in the
SETUP
list, using the
<
>
,
<
>
(or
<
>
,
<
>
) push-buttons
and press the
<ENTER>
.
The view of the display in the
SETUP
list with the RMS INTEGRATION text highlighted (displayed
inversely)
Two options are available:
LINEAR
and
EXPONENTIAL
. The required parameter can be selected
by means of the
<
>
,
<
>
push-buttons. The position is closed and the instrument returns to the
SETUP
list after pressing the
<ENTER>
(with the confirmation of a change made in the position) or the
<ESC>
push-button (ignoring a change made in the position).
The expressions used for the
LEQ
or
RMS
calculations are given in Appendix D. Setting
LINEAR
is
required for the IEC 61672-1 standard (for getting the true RMS value of the measured signal). When this
option is selected in the case of sound measurements, the value of the
LEQ
and
SEL
function (the SLM
case) does not depend on the detector time constant (the results are displayed
without
the indicator of
the detectors selected in the profiles).
The view of the displays with the available options of the RMS INTEGRATION position
Setting
EXPONENTIAL
enables the user to fulfil the requirements of the IEC 804 standard for the
LEQ
measurements. When this option is selected in the case of sound measurements, the value of the
LEQ
and
SEL
function (the SLM case) depends on the detector time constant (the results are displayed
with
the indicator of the detectors selected in the profiles).
In the case of sound measurements, setting
EXPONENTIAL
enables the user to fulfil the
requirements of the IEC 804 standard for the
RMS
measurements. When this option is selected the value
of the function depends on the detector time constant (the results are displayed
with
the indicator of the
detectors selected in the profiles).
Setting the reference signal in vibration measurements - REFERENCE LEVEL
The
REFERENCE LEVEL
(path: MENU / SETUP / REFERENCE LEVEL)
sub-list enables the user to
set the reference level of the signal in vibration or sound measurements. The values which are set here
Summary of Contents for SVAN 948
Page 1: ...SVAN 948 SOUND AND VIBRATION DATA LOGGER USER S MANUAL SVANTEK Sp z o o WARSAW June 2004...
Page 19: ...SVAN 948 USER MANUAL 2 6 The view of the rear panel of the SVAN 948 instrument in 1 1 scale...
Page 25: ...SVAN 948 USER MANUAL 3 6 Control diagram of the INPUT list...
Page 28: ...SVAN 948 USER MANUAL 3 9 Control diagram of the DISPLAY list...
Page 30: ...SVAN 948 USER MANUAL 3 11 Control diagram of the FILE list...
Page 32: ...SVAN 948 USER MANUAL 3 13 Control diagram of the AUX FUNCTION list...
Page 36: ...SVAN 948 USER MANUAL 3 17...
Page 37: ...SVAN 948 USER MANUAL 3 18 Control diagram of the SETUP list...