The user can select the following results, which can be saved:
PEAK, P–P, MAX
and
RMS
in the
case of vibration measurement. The selection of the parameter from the
CHANNEL x
sub-list is made by
pressing the
<
>
,
<
>
push-buttons. After pressing the
<ENTER>
push-button any changes made in the
sub-list are confirmed and the sub-list is closed. The return to the
BUFFERS SETUP
sub-list ignoring any
changes made in the sub-list is made after pressing the
<ESC>
push-button.
In order to switch parameters in the buffer, the user has to place special character in the chosen
position; available values are:
[
√
√√
√
]
or
[ ]
. The selection of the required buffers parameter is made with the
<
>
,
<
>
push-buttons. The user can pass to the selection of the next parameter in the
CHANNEL x
sub-list pressing the
<
>
,
<
>
push-buttons. After pressing the
<ENTER>
push-button any changes
made in the sub-list are confirmed and the sub-list is closed. The return to the
BUFFERS SETUP
sub-list
ignoring any changes made in the sub-list is made after pressing the
<ESC>
push-button.
The view of the displays in the CHANNEL 1 sub-list; the selection of the parameters
to be saved in the buffer’s file
The view of the displays in the BUFFERS SETUP sub-list; the all the parameters were selected
to be saved in the buffer’s file; in CHANNEL 1 and 2 vibration and in CHANNEL 3 and 4 sound parameters
Notice: The change of the channels parameters is not possible when the measurement is
performed.
The user has to finish the current measurement.
Notice:
Setting
BUFFER:OFF
results in bypassing the storing of the given channel’s results
in the buffer’s file.
It creates more space in the buffer for other channels
(see Chapter 5 for details).
Notice:
The buffer is updated in the time periods (steps) defined in the
BUF. STEP
(path:
MENU / INPUT / MEASURE SETUP / BUF. STEP)
. The way the measurement results are stored is described
in Chapter 5.
Selection of 1/1 OCTAVE analysis parameters for vibration signal - SPECTRUM
The
SPECTRUM
sub-list
(
path: MENU / INPUT / 1/1 OCTAVE SETUP / CHANNEL x / SPECTRUM
)
is a
context element, which appears on the
INPUT
sub-list in
1/1 OCTAVE
or
1/3 OCTAVE
mode. Using the
SPECTRUM
sub-list one can select in
1/1 OCTAVE
analysis the result which has to be saved in the
buffer’s file of the instrument.
Summary of Contents for SVAN 948
Page 1: ...SVAN 948 SOUND AND VIBRATION DATA LOGGER USER S MANUAL SVANTEK Sp z o o WARSAW June 2004...
Page 19: ...SVAN 948 USER MANUAL 2 6 The view of the rear panel of the SVAN 948 instrument in 1 1 scale...
Page 25: ...SVAN 948 USER MANUAL 3 6 Control diagram of the INPUT list...
Page 28: ...SVAN 948 USER MANUAL 3 9 Control diagram of the DISPLAY list...
Page 30: ...SVAN 948 USER MANUAL 3 11 Control diagram of the FILE list...
Page 32: ...SVAN 948 USER MANUAL 3 13 Control diagram of the AUX FUNCTION list...
Page 36: ...SVAN 948 USER MANUAL 3 17...
Page 37: ...SVAN 948 USER MANUAL 3 18 Control diagram of the SETUP list...