LAMBDA 10-3 OPERATION MANUAL – REV. 3.03 (20110829)
73
8.
8.
8.
8.
EXTERNAL PARALLEL INTERFACE CONTROL
EXTERNAL PARALLEL INTERFACE CONTROL
EXTERNAL PARALLEL INTERFACE CONTROL
EXTERNAL PARALLEL INTERFACE CONTROL
In Lambda 10-3’s programmed with firmware version 1.21 or later, the parallel port is
enabled for use.
When configured to do so, the Lambda 10-3 microprocessor sets the PARALLEL port as the
input source upon startup and anytime the ON LINE key is pressed. An appropriate
command from the remote computer through the PARALLEL port will also set that port as
the active input source and the display on the controller will reflect that change on its bottom
line. The PARALLEL port will remain active until an appropriate command is sent from a
remote computer through the SERIAL port or until the LOCAL key is pressed on the
controller keypad. This chapter discusses these connections and the structure of the data
that is required from your computer for controlling the filter wheel(s) and shutter(s) through
the PARALLEL port. Many of the details about the command structure that apply to use of
the SERIAL and USB ports, also apply to the PARALLEL port.
8.1
8.1
8.1
8.1
Connecting to the Parallel Port Interface
Connecting to the Parallel Port Interface
Connecting to the Parallel Port Interface
Connecting to the Parallel Port Interface
The parallel interface connector on the back of the cabinet is a DB-25 connector, as shown in
the following figure.
Parallel Port
Figure 8-1. Location of parallel port on rear of Lambda 10-3 controller cabinet.
This port is designed for connection to a standard parallel port on a computer and is not to be
connected to any other type of device.
The pin assignments for this connector are, as viewed from the back of the unit, shown
below.
Figure 8-2. Parallel interface DB-25 connector on the Lambda 10-3.
Summary of Contents for Lambda 10-3
Page 12: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 xii This page intentionally blank ...
Page 46: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 34 This page intentionally blank ...
Page 72: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 60 This page intentionally blank ...
Page 84: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 72 This page intentionally blank ...
Page 96: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 84 This page intentionally blank ...
Page 102: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 90 This page intentionally blank ...
Page 106: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 94 This page intentionally blank ...
Page 108: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 96 This page intentionally blank ...
Page 112: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 100 This page intentionally blank ...
Page 128: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 116 This page intentionally blank ...
Page 132: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 120 NOTES NOTES NOTES NOTES ...