LAMBDA 10-3 OPERATION MANUAL – REV. 3.03 (20110829)
50
Command
Command
Command
Command
Value
Value
Value
Value
(Decimal,
hexadecimal, &
binary)
Description
Description
Description
Description
Close Shutter A
Close Shutter A
Close Shutter A
Close Shutter A
172
AC
10101100
Sets the state of Shutter A closed.
Open Shutter B
Open Shutter B
Open Shutter B
Open Shutter B
186
BA
10111010
Sets the state of Shutter B to open.
Open Shutter B
Open Shutter B
Open Shutter B
Open Shutter B
Conditionally
Conditionally
Conditionally
Conditionally
187
BB
10111011
Configures Shutter B to open while the filter wheel is
stopped. The shutter will close each time the filter wheel
moves and then will open again at the end of the move.
Close Shutter B
Close Shutter B
Close Shutter B
Close Shutter B
188
BC
10111100
Sets the state of Shutter B closed.
Open Shutter C
Open Shutter C
Open Shutter C
Open Shutter C
234
EA
11101010
Sets the state of Shutter C to open.
Open Shutter C
Open Shutter C
Open Shutter C
Open Shutter C
Conditionally
Conditionally
Conditionally
Conditionally
235
EB
11101011
Configures Shutter C to open while the filter wheel is
stopped. The shutter will close each time the filter wheel
moves and then will open again at the end of the move.
Close Shutter C
Close Shutter C
Close Shutter C
Close Shutter C
236
EC
11101100
Sets the state of Shutter C closed.
Fast
Fast
Fast
Fast----mode Shutter
mode Shutter
mode Shutter
mode Shutter
220 + 1-3
DC + 01-03
11
00000001-
00000011
Sets the shutter to fast mode (SmartShutter only).
Second byte indicates Shutter A (1), B (2), or C (3).
Soft
Soft
Soft
Soft----mode Shutter
mode Shutter
mode Shutter
mode Shutter
221 + 1-3
DD + 01-03
10
00000001-
00000011
Sets the shutter to soft mode (SmartShutter only).
Second byte indicates Shutter A (1), B (2), or C (3).
Neutral Density
Neutral Density
Neutral Density
Neutral Density----
mode Shutter
mode Shutter
mode Shutter
mode Shutter
222 + 1-3 +
1 – 144
DE + 01-03 +
01 – 90
10
00000001-
0
00000001 -
10010000
Sets the shutter to neutral-density mode (SmartShutter
only). Second byte indicates Shutter A (1), B (2), or C (3).
The third byte contains a value of 1 through 144
indicating the number of microsteps.
Summary of Contents for Lambda 10-3
Page 12: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 xii This page intentionally blank ...
Page 46: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 34 This page intentionally blank ...
Page 72: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 60 This page intentionally blank ...
Page 84: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 72 This page intentionally blank ...
Page 96: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 84 This page intentionally blank ...
Page 102: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 90 This page intentionally blank ...
Page 106: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 94 This page intentionally blank ...
Page 108: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 96 This page intentionally blank ...
Page 112: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 100 This page intentionally blank ...
Page 128: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 116 This page intentionally blank ...
Page 132: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 120 NOTES NOTES NOTES NOTES ...