Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the single-
character address
XT Set
Averaging
Time
(
See XPA
command
)
aXT+t!
t = averaging time in seconds (0 to
240 seconds)
Example: 0XT+10!
(sets the averaging time to 10
seconds if bubbler operation
disabled)
a0011
indicating the command will be
complete in 1 second and 1 value can be
collected
Note:
a D0 command issued after the
XT will return the number of samples to
be averaged during the selected time.
XUP Set pressure
units
aXUP+n+d!
n = 0
feet water
n = 1
psi
n = 2
kPa
n = 3
cm water
n = 4
meters water
n = 5
mm water
n = 9
user units
d = number of places right of the
decimal (optional parameter)
Example: 0XUP+9+2!
select user units with 2 right digits--
(make sure you use XUU command
to set the scale and offset for the
desired user units)
a0012
indicating the command will be
complete in 1 second and 2 values can
be collected
Note:
a D0 command issued after the
XUP will return the value of the units
that are selected and the number of
digits right of the decimal point.
XUT Set
temperature
units
aXUTn!
n = 0 for Celsius,
n = 1 for Fahrenheit
Example: 0XUT1!
(set temperature units to F)
a0011
indicating the command will take 1
second to complete and 1 value can be
collected.
Note:
a D0 command issued after the
XUT will return the value of the units
that are selected.
XU
U
Set User
Units
aXUUso!
where s is the pressure scale factor
and o is the offset,
User output = (psi)*scale + offset
Example: 0XUU+27.63+0
(27.63 inches per psi)
a0012
Note:
a D0 command issued after XUU
will return the scale and offset.
Note
: a scale of 0 is invalid.
Note:
Be sure that the units of pressure
(XUP) are set to user units (9).
40