Sutron AccuBubble Extended Commands
Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the
single-character address
X? Request
unknown
address
*X?!
This command causes any Sutron
ACCUBUBBLE to identify itself. If
you have more than one
ACCUBUBBLE connected, the
result may be garbled. There is no
guarantee that non-ACCUBUBBLE
devices will respond to this
command.
a
ACCUBUBBLE’s address.
XA
D
Set SDI-12
address
aXADnAn!
n new SDI-12 address, repeated
twice
Example: 5XAD9A9!
(set address 5 to address 9, the
address was previously set to 5)
a0011
no response if addresses do not match
Note:
if the DIP switches are set to a
non-zero address then upon power-up
the address will be the dip-switch
address.
Note:
a D0 command issued to the new
address after the XAD command will
return the new address.
XE Set Field
Calibration
offset
aXExu!
where x is the pressure offset
u indicates the units of the offset,
0 = feet water, 1 = psi.
Example: 0XE-0.05+0
(set offset to -0.05 feet)
a0011
indicating that the command will take 1
second and 1 value can be collected.
a
service request
Note:
a D0 command issued after the
XE command will return the offset in
units of psi for software versions before
2.0. Version 2.0 and later return the
offset in the current units of pressure.
XFD Reset
to
Factory
defaults
aXFD!
This command resets most user
configurable configuration items
back to the factory defaults. It does
not reset the address nor does it
affect metrology lab calibrations.
Note:
It is recommended that the unit
be powered down and back up after
use of this command.
a0101
indicating that the command will take 10
seconds and 1 value can be collected.
a
service request
Note:
a D0 command issued after the
XFD command will return the operating
mode
Bringing the Benefits of Real-Time Data Collection to the World