Command
Description
Command Syntax
(command underlined)
Sensor response (underlined)
"a" represents the
single-character address
C7 Request
Concurrent
factory
calibration
Pressure and
Temperature
Measurement
(Version 2.0
and higher)
aC7!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready and 02 is the
number of values that can be collected
apt
where p is the pressure psi and t is the
temperature in degrees Celsius.
CC Request
Default
Concurrent
Pressure
Measurement
with CRC-16
(Version 2.0
and higher)
aCC!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready, 2 is the number
of values that can be collected
axuC
where x is the signed pressure value, u is
the signed indicator of the units, and C
is the 3 character CRC. The units are set
by the XUP command.
CC1 Request
Concurrent
Pressure
Measurement
in psi with
CRC-16
(Version 2.0
and higher)
aCC1!
aD0!
attt01
ttt is the time in seconds until the
measurement is ready, 1 is the number
of values that can be collected.
apC
where p is the signed pressure value in
psi and C is the 3 character CRC
CC2 Request
Concurrent
Temperature
Measurement
with CRC-16
(Version 2.0
and higher)
aCC2!
aD0!
attt02
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
atuC
where t is the temperature, u is the units
0= Celsius and 1=Fahrenheit, and C is
the 3 character CRC. Use the XUT
command to set the units.
CC3 Request
User
Scale, User
Offset, and
Field
Calibration
Offset with
CRC-16
(Version 2.0
and higher)
aCC3!
aD0!
a00003
000 is the time in seconds until the
measurement is ready and 3 is the
number of values that can be collected
asocC
where, s is the user scale and o is the
user offset (psi), c is the field calibration
offset in the current pressure units, and
C is the 3 character CRC.