Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the
single-character address
MC Request
Default
Pressure
Measurement
with CRC-16
(Version 2.0
and higher)
aMC!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready, 2 is the number
of values that can be collected
a
service request
axuC
where x is the signed pressure value, u is
the signed indicator of the units, and C
is the 3 character CRC. The units are set
by the XUP command.
MC1 Request
Pressure
Measurement
in psi with
CRC-16
(Version 2.0
and higher)
aMC1!
aD0!
attt1
ttt is the time in seconds until the
measurement is ready, 1 is the number
of values that can be collected.
a
service request
apC
where p is the signed pressure value in
psi and C is the 3 character CRC
MC2 Request
Temperature
Measurement
with CRC-16
(Version 2.0
and higher)
aMC2!
aD0!
attt2
ttt is the time in seconds until the
measurement is ready and 2 is the
number of values that can be collected
a
service request
atuC
where t is the temperature, u is the units
0= Celsius and 1=Fahrenheit, and C is
the 3 character CRC. Use the XUT
command to set the units.
MC3 Request
User
Scale, User
Offset, and
Field
Calibration
Offset with
CRC-16
(Version 2.0
and higher)
aMC3!
aD0!
a0003
000 is the time in seconds until the
measurement is ready and 3 is the
number of values that can be collected
asocC
where, s is the user scale and o is the
user offset (psi), c is the field calibration
offset (current pressure units), and C is
the 3 character CRC.